Supporting Research
  Is the Otogram™ As Accurate As Manual Testing?
“An audiologist would have to see a change in test-retest audiometric thresholds of approximately 10-15 dB to be 95% confident that the difference did not occur as a result of variability due to measurement error.” [Stuart, Stenstrom, Tompkins, Vandenfoff, 1991]

Tympany, with many of its customers, have tested hundreds of patients in side-by-side tests to compare automated Otogram™ audiogram testing results to conventional manual audiogram testing results. These comparisons have consistently shown that the test results obtained with the Otogram™ vary to the same degree as the test-retest difference found in manual testing – that is, nearly all the thresholds measured in manual and automated Otogram™ testing have fallen within 10 dB of one another.


Clinical Studies of Test-Retest Variability
Published studies of test-retest variability indicate that in measuring air conduction thresholds, audiologists would expect a 10 dB variance, at a 95% confidence interval, between thresholds for the same patient when measured at two different points in time by two different audiologists. [Stuart , Stenstrom , Tompkins, Vandenhoff 1991] In addition, research on automated pure-tone air conduction audiometry by the VA has indicated that when comparing manual to automated audiometry, 98% of the thresholds were within 10 dB, with 100% within 20 dB. [Henry, Flick, Gilbert, Ellingston, Fausti 2001]


Manual versus Otogram™ Testing Protocols
Nearly all side-by-side testing was completed using the following methodology. A patient was taken into a sound booth and given a comprehensive audiogram by an audiologist. The same patient was then taken to a quiet examination room and prepared by a testing administrator to self administer the Otogram™ test. All the testing results were then compared to see the variance between manual and Otogram™ results.


Testing Population
Side-by-side testing was done in several clinical locations in Texas and Virginia by audiologists. In total, 753 thresholds were measured and compared for 43 patients.

Summary Testing Results

Summary of Results
94% of all thresholds fel within 10 db of manual results
97% of all air thresholds fell within 10 dB of manual results
84% of all bone thresholds fell within 10 dB of manual results
98% of all SRT thresholds fell within 10 dB of manual results
Average difference in speech discrimination, 3.5%

Pure Tone and SRT Results
Total
Air
Bone
SRT
Thresholds measured
753
795
198
82
0 dB difference to manual
39%
42%
29%
46%
0-5 dB difference to manual
79%
86%
60%
79%
0-10 dB difference to manual
94%
97%
84%
98%
0-15 dB difference to manual
99%
100%
97%
100%
Over 15 dB difference to manual
1%
0%
3%
0%

Testing Time
Total
Air
Bone
SRT
Time to complete tests
(avg min)
20.4
15.7
2.8
1.9

We currently have partnered with several academic institutions and researchers to conduct additional studies and will publish results upon completion. If you are interested in conducting research pertaining to the use of the Otogram, please contact us directly.